Kewaunee K1
Fume Hood · S/N 34895
Recurring issue – 3+ repairs in 6 months Renew Contract
Equipment Information
Serial No.
34895
Brand
Kewaunee
Model
K1
Customer
Tan Tock Seng Hospital
Department
Clinical Research Unit
Install Date
20 Nov 2022
Person in Charge
Mr. Goh Kian Seng
Email
kianseng_goh@ttsh.com.sg
Phone
+65 6357 7000
Warranty
Expired20 Nov 202219 Nov 2024
Expired 701 days ago
Service Contract
Expiring01 Dec 2024 – 30 Nov 2026
PM 1 of 3 visits completed
Next: 30 Oct 2026
Service History
- Service Case 187 · Waiting for P.O.25 Sept 2026Engineer: Jason LeeView Report →Problem: Exhaust fan noiseAction taken: Bearings lubricated; fan motor replacement quoted
- Service Case 158 · Completed19 Aug 2026Engineer: Ravi KumarView Report →Problem: Airflow alarm recurringAction taken: Replaced controller board; monitoring
- Service Case 142 · Completed03 Jul 2026Engineer: Daniel TanView Report →Problem: Sash not holding positionAction taken: Replaced sash counterweight cable
- PM Visit · Completed20 May 2026Engineer: Siti NurhalizaView Report →Problem: Inspection – PM visit 1 of 3 (AGR-2024-031)Action taken: Inspection completed, all parameters within spec
- Service Case 121 · Completed12 May 2026Engineer: Ravi KumarView Report →Problem: Airflow alarm triggering intermittentlyAction taken: Replaced airflow sensor, recalibrated face velocity
- Installation · Completed20 Nov 2022Engineer: Ravi KumarView Report →Problem: New installation & commissioningAction taken: Installed, commissioned and user training conducted